- SP Scientific VirTis 50L Ultra EL-85°C Pilot Clean Room Configuration Freeze Dryer Lyophilizer (100012485)
- UPS Battery Backup
- Computer (with Monitor, Keyboard, Mouse) and LyoS 2.0 Control Software
- Communication Cable
- Attached Power Cable
- 90-Day Warranty
Notice: This freeze dryer is water cooled. You will need to meet this utility requirement to operate this freeze dryer.
After its initial disinfection and inspection, our technicians tested the SP Scientific VirTis 50L Ultra EL-85°C Pilot-Production clean room freeze dryer with the following results:
- Condenser temperature achieved -84.1°C: PASSED
- Shelf temperature achieved -72.4°C: PASSED
- Vacuum achieved 16mT: PASSED
Manufacturer
|
SP Scientific
|
Model
|
50L Ultra EL-85°
|
Part No.
|
100012485
|
Serial No.
|
322291
|
Lowest shelf temperature (50Hz / 60Hz)
|
≤ -67°C / -70°C
|
Shelf temperature control range
|
-55 to 65°C
|
Shelf pull-down from 20°C to -40°C
|
≤ 40 minutes
|
Lowest condenser temperature (50Hz / 60 Hz)
|
≤ -82°C / -85°C
|
Maximum condenser capacity
|
≥ 50 L
|
Condenser surface area
|
10 ft2 (.93 m2)
|
Condenser pull-down from 20°C to -45°C
|
≤ 35 minutes
|
Maximum ice condensing capacity in 24 hours
|
≥ 20 L
|
Maximum deposition rate
|
≥ 0.83 L/hour
|
Number of compressors
|
2
|
Compressor horsepower
|
3.5 hp, 1.5 hp
|
Vacuum time to 100 mTorr
|
≤ 45 minutes
|
Vacuum rate of rise
|
≤ 30 mT/hour
|
Vacuum control +/- 5 mTorr
|
50mTorr to 500mTorr
|
Volume-based leak rate
|
≤ .0042 mbar L/sec
|
Lowest system vacuum
|
≤ 15 mT
|
Temperature uniformity
|
± 1.0°C
|
Construction
|
316L stainless steel shelves, product chamber and condenser chamber
|
Stoppering
|
Bottom-up hydraulic
|
Defrost type
|
Hot gas
|
Refrigerant type
|
CFC-free
|
Vapor port
|
8 in (20.3 cm)
|
Power rating
|
208-230VAC, 60Hz, 50A, 1Ph
220VAC, 50Hz
|
Recommended outlet
|
NEMA 6-50R
|
Overall dimensions (W x D x H)
|
25 x 78 x 75.75 in (64 x 198 x 192 cm)
|
Minimum clearance
|
10 in (25.4 cm)
|
Weight
|
2,000 lb (909 kg)
|
Note: Performance specifications are based on SP test data from units operating at an ambient room temperature of approximately 20°C. SP recommends an optimum operation range of 15-25°C (59-77°F).
Note: SP recommends a 24-inch (61 cm) clearance around all sides of the unit for serviceability. If machines are placed side by side, increase the minimum clearance to 48 inches (121.9 cm)
|
Water-Cooled
|
Compressed air (for units with isolation valve)
|
80 psig (5.5 bar)
|
Ambient room temperature
|
15-25°C (59-77°F)
|
Approx. peak room heat generated
|
5,400 BTU/h (1.6 kW)
|
Cooling water usage
|
2-5 gpm (8-19 Lpm)
|
Cooling water load
|
17,500 BTU/h (5.1 kW)
|
Electrical power rating
|
208/230VAC, 60Hz, 50A, 1Ph
|
Recommended power outlet
|
NEMA 6-50R
|
|
Gas 1
|
Gas 2
|
F gas
|
R407C
|
R508B
|
Charge (Kg)
|
1.245
|
0.970 (AC); 0.452 (WC)
|
GWP
|
1774
|
13396
|
EPA SNAP
|
IP/VLTR
|
IPR/VLTR
|
Safety Class
|
A1
|
A1
|
Total CO2e (t)
|
15.198 (AC); 8.264 (WC)
|
15.198 (AC); 8.264 (WC)
|