- Tecan Infinite M200 Pro Microplate Reader (30050303)
- Dell OptiPlex 5080 Computer with Magellan v7.4 Installed
- Monitor, Mouse, and Keyboard
- USB Cable
- Display Cable
- x3 Power Cables
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Tecan Infinite M200 Pro microplate reader with the following results:
- Powers on and initializes: PASSED
- On board plate load button: PASSED
- Multiple test assays performed with consistent results: PASSED
Note: It is possible that some of these specifications do not apply to this item, specifically. They are covering the entire Infinite M200 line, not just the M Nano+ configuration.
Manufacturer
|
Tecan
|
Model
|
Infinite 200 Pro
|
Configuration
|
Infinite M200 Pro
|
Reference No.
|
30050303
|
Serial No.
|
1211001078
|
Mfg. date
|
20121114
|
Power rating
|
100-120VAC / 220-240VAC, 50/60Hz, 150VA
|
Dimensions (W x D x H)
|
17 x 18 x 9.5 in
|
Light source
|
UV Xenon flashlamp
|
Wavelength selection
|
Quad4 Monochromators system (2 excitation and 2 emission monochromators)
|
Bandwidth
|
Ex: < 5 nm for λ ≤ 315 nm and < 9 nm for λ > 315 nm; Em: < 20 nm
|
Wavelength accuracy
|
Absorbance: < ± 0.5 nm for λ > 315 nm; < ± 0.3 nm for λ ≤ 315 nm Fluorescence: < ± 2 nm for λ > 315 nm; < ± 1 nm for λ ≤ 315 nm
|
Wavelength reproducibility
|
Absorbance: < ± 0.5 nm for λ > 315 nm; < ± 0.3 nm for λ ≤ 315 nm Fluorescence: < ± 1 nm for λ > 315 nm; < ± 0.5 nm for λ ≤ 315 nm
|
Wavelength range
|
Fluorescence intensity: Ex 230 – 850 nm, Em 280 – 850 nm Absorbance: 230 - 1000 nm
|
Detectors
|
Fluorescence – PMT, UV and red-sensitive Absorbance – UV silicon photodiode Luminescence –photon counting system with low dark current PMT
|
Plate formats
|
6- to 384-well plates, cuvettes, NanoQuant plate
|
Temperature control
|
Ambient +5°C up to 42°C
|
Shaking
|
Linear, orbital
|
Fluorescence sensitivity values
|
Top reading: 170 amol / well (100 μl; 384-well plate) Bottom reading: 9.0 fmol / well (200 μl; 96-well plate) TRF: 90 amol / well (100 μl; 384-well plate)
|
Absorbance ratio accuracy @ 260 / 280 nm
|
± 0.07
|
Precision @ 260 nm
|
< 0.2%
|
Accuracy @ 260 nm
|
< 0.5%
|
Measurement range
|
0 - 4 OD
|
Injector pump speed
|
100 - 300µl/s
|
Injection volume
|
selectable in 1µl increments; max. volume: 800 µl per stroke
|
Dead volume
|
100 µl including pump back
|
Fastest read times
|
96-well plate: 20 sec 384-well plate: 30 sec Wavelength Ex / Em-scan, 96-well plate (450 - 550 nm, 5 nm step): 150 sec
|