- Thermo Microm HM 550 OVP Cryostat Microtome
- Option V - Vacuum Generated Unit for stretching sections and disposing of section waste
- Option P - Active Deep Freezing Station with Peltier Element
After its initial disinfection and inspection, our technicians tested the HM 550 OVP with the following results:
- Cooling head reaches a -50°C temperature: PASSED
- Compressors function properly: PASSED
- Condenser fan functions properly: PASSED
- Door seal is in good condition: PASSED
- Control panel functions properly: PASSED
- Temperature sensors function properly: PASSED
- Unit reaches setpoint and maintains temperature reliably and reached -35°C: PASSED
- Door security function is working properly: PASSED
- Maintains manufacture temperature specs: PASSED
- 2nd stage compressor oil flush
- Replaced/reinforced insulation
|
|
|
|
|
|
|
|
|
|
Chamber temperature control
|
|
Specimen temperature control
|
|
Fine sectioning thickness
|
|
Trim sectioning thickness
|
|
|
|
|
|
Fast freezing station depending on the chamber temperature
|
|
Actively cooled fast freezing station with Peltier element
|
|
|
automatic defrosting, adjustable with temperature control
manual immediate defrosting
|
|
|
|
|
|
|
Specimen retraction during return travel
|
|
|
|
Horizontal specimen movement
|
|
|
Alternatively manual with variable speed
|
|
special sizes upon request to manufacturer
|
|
universal 8° with zero positioning
|
|
motorized, graduated, single and continuous
|
|
|
Cooling chamber illumination
|
with variable illumination position
|
Dimensions (w/o handwheel) (W x D x H)
|
30” x 25” x 47” (64 x 76 x 120 cm)
|
|
|