- Thermo Scientific Microm CryoStar NX50 HOPD Cryostat Microtome (957110)
- Knife/Blade Holder Assembly
- x3 Specimen Discs
- Power Cable
- 90-Day Warranty
Option H - Height adjustment that raises and lowers the work area
Option O - Specimen cooling
Option P - Peltier cooling cryobar
Option D - Cold disinfection
After its initial disinfection and inspection, our technicians tested the Thermo Scientific Microm CryoStar NX50 HOPD cryostat microtome with the following results:
- Specimen cooling maintains -40°C setpoint (manufacturer’s spec): PASSED
- Height adjustment: PASSED
- Peltier cooling: PASSED
- Replaced fan motor.
- Installed a used nebulizer.
- Drained, flushed, and filled the tyfoxit cooling unit.
Manufacturer
|
Microm
|
Model
|
CryoStar NX50 HOPD
|
Ref. No.
|
957110
|
Serial No.
|
60824
|
Fine section thickness range
|
0.5 - 100 µm
|
Trimming thickness range
|
5 - 500 µm
|
Vertical specimen stroke
|
64 mm (2.5 in)
|
Horizontal knife movement
|
48 mm (1.9 in)
|
Specimen orientation
|
x- and y- axes universal 8° with zero positioning; z-axis up to 360°
|
Specimen retraction
|
20 µm
|
Specimen temperature control
|
-43°C ±2°C (-45.4°F) at 20°C (68°F) ambient
|
Cryobar cooling
|
-57°C ±3°C (-70.6°F) at 20°C (68°F) ambient
|
Cryochamber
|
-25°C ±2°C (-13°F) at 20°C (68°F) ambient; continuously cooled
|
Knife holder
|
-27°C ±2°C (-16.6°F) at 20°C to 30°C (68°F to 86°F) ambient; continuously cooled
|
Defrosting options
|
Programmed / Immediate
|
Certifications
|
CSA, CE
|
Electrical requirements
|
110-120V~, 60Hz, 10A
|
Dimensions (W x D x H)
|
33 x 38 x 43 to 54.5 in
|
Weight
|
441 lb (200 kg)
|