- Thermo Scientific NanoDrop 8000 UV-Vis Spectrophotometer (ND8000)
- Dell lLaptop with NanoDrop 8000 v2.3.3 Software installed
- USB Cable
- NanoDrop Power Supply
- Laptop Power Supply
- x2 Power Cables
- 90-Day Warranty
 
After its initial disinfection and inspection, our technicians tested the Thermo Scientific NanoDrop 8000 UV-Vis spectrophotometer with the following results: 
 
- Ran diagnostics check: PASSED
- Ran intensity check: PASSED
- Ran spectrum scan of various liquids with peaks in appropriate wavelengths: PASSED
 
| Manufacturer | Thermo Scientific | 
| Product line | Nanodrop | 
| Model | ND8000 | 
| Serial No. | 0990 | 
| Absorbance accuracy | 3% (at 0.74 Abs at 350 nm) | 
| Absorbance range | 0 to 75 A | 
| Certifications/compliance | UL/CSA and CE | 
| Compatibility | Professional versions of Microsoft Windows 7 (32 bit and 64 bit), Windows Vista (32 bit), and Windows XP (32 bit) | 
| Connections | USB | 
| Detection limits | 2.5 ng/uL (dsDNA), 0.15 mg/mL (BSA) | 
| Detection range | 2.5 to 3,700 ng/uLL (dsDNA), 0.15 - 100 mg/mL (BSA) | 
| Detector type | 2048-element linear silicon CCD array | 
| Light source | Xenon flash | 
| Type | Microvolume UV-Vis Spectrophotometer | 
| Wavelength range | 220 to 750 nm | 
| Footprint | 24 x 32 cm | 
| For use with (application) | Nucleic acid, protein, cell culture and custom methods | 
| Measurement time | <20 sec. | 
| No. of samples | 8 | 
| Pathlength | 0.2 and 1.0 mm | 
| Sample volume | 1uL | 
| Spectral resolution | ≤3 nm (FWHM at Hg 546.1 nm) | 
| Wavelength accuracy | +/-1 nm | 
| Power input | 12VDC, 30W | 
| Dimensions (W x D x H) | 9 x 12.6 x 7 in | 
| Weight | 7.5 lbs (3.4 kg) |