- Thermo Scientific Shandon CryoStar NX70 HOMVPD Cryostat Microtome (957030)
- Knife/Blade Holder Assembly with Glass Anti-Roll Plate
- x2 Specimen Discs
- Heat Extractor
- Magnetic Blade Remover
- Power Cable
- 90-Day Warranty
Option H - Height adjustment that raises and lowers the work area
Option O - Specimen cooling
Option M - Motorized cutting
Option V - Vacutome
Option P - Peltier cooling cryobar
Option D - Cold disinfection
After its initial disinfection and inspection, our technicians tested the Thermo Scientific Shandon CryoStar NX70 HOMVPD cryostat microtome with the following results:
- Chamber cooling -35°C setpoint: PASSED
- Specimen cooling -40°C setpoint: PASSED
- Knife cooling -35°C setpoint: PASSED
- Motorized cutting drive: PASSED
- Height adjustment: PASSED
- Peltier cooling: PASSED
- Vacutome: PASSED
- Checked sample advance: PASSED
- Checked chamber light: PASSED
- Checked disinfection system: PASSED
- Checked window heater: PASSED
- Verify e-Stop: PASSED
- Verify vacuum assist: PASSED
- Verify handwheel brake: PASSED
- Check fan alignment and service: PASSED
- Replaced vacuum filter.
- Replaced disinfection nebulizer.
- Flushed and filled tyfoxit fluid for specimen cooling.
- Cleaned undercarriage.
- Cleaned refrigeration condenser.
- Cleaned chamber, microtome, and knife assembly.
- Lubricated mirotome and knife assembly.
Manufacturer
|
Thermo Fisher Scientific
|
Model
|
CryoStar NX70 HOMVPD
|
Cat. No.
|
957030
|
Serial No.
|
S18071354
|
Section thickness range
|
0.5 to 500µm
|
Options Type
|
HOMVPD
|
Fine section thickness range
|
0.5 to 100µm
|
Fine section thickness resolution
|
0.5 to 5µm in 0.5µm steps, 5 to 10µm in 1µm steps, 10 to 20µm in 2µm steps, 20 to 50µm in 5µm steps, 50 to 100µm in 10µm steps
|
Trim thickness range
|
5 to 500µm
|
Trim thickness resolution
|
0.5 to 5µm in 0.5µm steps, 5 to 10µm in 1µm steps, 10 to 20µm in 2µm steps, 20 to 50µm in 5µm steps, 50 to 100µm in 10µm steps
|
Vertical cutting stroke
|
64mm
|
Horizontal specimen movement
|
48mm
|
Specimen orientation
|
x- and y-axes universal 7° with zero positioning; z-axis up to 360°
|
Cutting drive
|
Manual and motorized, electronically controlled
|
Cutting speed
|
0 to 256mm/sec.
|
Cutting window
|
Manually adjusted to specimen size
|
Chamber temperature
|
-60°C max.
|
Specimen temperature control
|
-50° to +10°C (-58° to +50°F) at +20°C room temp. (decrease 0.7°C/1°C room temp. increase)
|
Knife temperature control
|
-35° to -5°C (-31° to +23°F) at 20°C room temp. (decrease 0.5°C/1°C room temp. increase)
|
Display
|
Integrated color LCD touch screen
|
Specimen retraction
|
20µm
|
Lighting
|
LED
|
Coarse feed
|
Motorized (on/off)
|
Clearance angle adjustment
|
Disposable blade carrier 8 to 16°
|
Defrost
|
Programmed/Immediate
|
Dimensions (W x D x H)
|
29.7 x 39.4 x 32.3 to 44.0 in (75.5 x 100 x 82 to 112 cm)
|
Weight
|
440 lb (200 kg)
|