Beckman Coulter Biomek NXp Automated Liquid Handling Workstation
 
Successful Software Communication & Plate Movement 
 
After connecting the NXp to the Biomek software, we performed the following tests:
 
    - NXp initializes and communicates with Biomek software: PASSED
    - Home all of the motors: PASSED
    - Prime syringe pumps: PASSED
    - Aspirate/dispense fluids: PASSED
    - Load/unload tips: PASSED
    - Lift/move microplates: PASSED
    - Create/edit/remove programs: PASSED
    - Beckman Coulter Biomek NXp Automated Liquid Handler
    - MultiChannel 96 P20/200 Head with Gripper
    - PC with software installed
    - Monitor
    - Mouse, keyboard
    - Tip loader
    - Heating block with controller
    - Accuframe sensor
    - Accuframe tool
    - Metal reservoir with adapter plate
    - Serial cable
    - Power cables
Consumables are required for this system to be used in production. You will need to acquire the consumables separately. They can generally be purchased from the manufacturer or a third-party dealer.
 
| 
 Manufacturer 
 | 
 Beckman Coulter 
 | 
| 
 Model 
 | 
 Biomek NXp 
 | 
| 
 Catalog No. 
 | 
 A31841 
 | 
| 
 Serial No. 
 | 
 A318410148 
 | 
| 
 Environment 
 | 
 Indoor use only 
 | 
| 
 Compressed Air Requirements 
NX-MC: Requires compressed air NX-S8: Only Span-8 instruments that use a Microplate Shaking ALP require compressed air 
 | 
 Air Pressure: 60 - 100 psi of compressed air 414 - 690 kPa of compressed air Air Flow: 2.25 cfm @ 60 psi 0.064 m3/min @ 414 kPa 
 | 
| 
 System Fluid Requirements 
NX-S8: Requires system fluid 
 | 
 De-ionized or Distilled water System fluid ambient operating temperature is 59°–86°F (15°–30°C). System fluid should be degassed for 24 hours prior to use. 
 | 
| 
 Weight 
 | 
 190 lbs. (86.2 kg) 
 | 
| 
 Ambient Operating Temperature 
 | 
 41°-86°F (5°-30°C) 
 | 
| 
 Humidity Restrictions 
 | 
 <85% (non-condensing) @ 86°F (30°C) 
 | 
| 
 Altitude Restrictions 
 | 
 up to 6562 ft (2000 m) 
 | 
| 
 Installation Category 
 | 
 Category II 
 | 
| 
 Pollution Degree 
 | 
 2 
 | 
| 
 Sound Pressure Level 
 | 
 Maximum sound pressure:  82 dB Maximum sound pressure at 1 meter away: 75 dB 
 | 
| 
 Fuses 
 | 
  US: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, UL recognized, CSA certified Europe: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, CENELEC approved 
 | 
| 
 Communications to Host 
 | 
 RS-232 port 
 | 
| 
 Communications to Tip Loader NX-MC: Uses a Tip Loader 
 | 
 CAN 
 | 
| 
 Communications to Active ALPs via the Biomex NXp Instrument 
 | 
 CAN 
 | 
| 
 Power Requirements 
 | 
 100-240V, 4A, 50/60Hz 
 | 
| 
 Overall Dimensions (WxDxH) 
 | 
 35 x 31 x 42" 
 |