- Beckman Coulter Vi-CELL XR Cell Viability Analyzer (383556)
- Computer with Vi-CELL XR 2.04 Software Installed
- Monitor, Keyboard, and Mouse
- Communication Cables
- Power Cables
- 90-Day Warranty
 
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Vi-CELL XR cell viability analyzer with the following results: 
 
- Powers on and initializes; connects to control software: PASSED
- Reagent pack error occurred as expected: PASSED
- Flushed lines with water to test reagent sensor: PASSED
- Pumps checked: PASSED
- Camera image checked: PASSED
 
Notice: The flow cell has been protected with a piece of cardboard to prevent the camera from smashing the cell during transit. It will need to be removed before use.
 
| 
 Manufacturer 
 | 
 Beckman Coulter 
 | 
| 
 Model 
 | 
 Vi-CELL XR 
 | 
| 
 Reference No. 
 | 
 383556 
 | 
| 
 Serial No. 
 | 
 AW09236 
 | 
| 
 Mfg. date 
 | 
 2014/02/28 
 | 
| 
 Autosampler 
 | 
 Yes 
 | 
| 
 Size range 
 | 
 2 - 70 um 
 | 
| 
 Sample volume 
 | 
 500 uL 
 | 
| 
 Analysis time 
 | 
 2.5 min 
 | 
| 
 Imaging technology 
 | 
 Camera focus: auto Image collection: Firewire Camera: 1394x1040 CCD 
 | 
| 
 Image zoom 
 | 
 Yes 
 | 
| 
 Aspiration and Trypan Blue mixing 
 | 
 Variable 
 | 
| 
 Concentration range 
 | 
 5 x 104 to 1 x 107 cells/mL Counting accuracy: +/- 6% 
 | 
| 
 Operating system 
 | 
 Windows 7 or Windows 10 
 | 
| 
 Instrument type 
 | 
 Video imaging through a quartz flow cell 
 | 
| 
 Temperature 
 | 
 10° to 40°C (50°F to 104°F) 
 | 
| 
 Power rating 
 | 
 100/240V, 50/60Hz, 0.5/1A 
 | 
| 
 Power consumption 
 | 
 120W 
 | 
| 
 Dimensions (W x D x H) 
 | 
 15 x 17.5 x 17.5 in 
 | 
| 
 Weight 
 | 
 36.2 lb (16.4 kg) 
 |