- Beckman Coulter Avanti J-25I High-Speed Refrigerated Floor Centrifuge (363106)
- Attached Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Avanti J-25I refrigerated floor centrifuge with the following results:
- Ran to 20,000 rpm: PASSED
- Refrigeration held at 4°C: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no issue found: PASSED
Manufacturer
|
Beckman Coulter, Inc.
|
Model
|
Avanti J-25I
|
Cat. No.
|
363106
|
Serial No.
|
JJY00H07
|
Mfg. date
|
2000
|
Speed range
|
100 to 25,000 rpm
|
Setting range
|
In 100-rpm increments below 10,000 rpm In 500-rpm increments above 10,000 rpm or equivalent rcf (in 100 x g increments)
|
Speed display
|
from 0 to 10,000 rpm, indicates actual rotor speed ±10 rpm from 10,000 to 25,000 rpm, indicates actual rotor speed ± 0.1%, or equivalent rcf
|
Time setting range
|
to 180 min or continuous (hold)
|
Time display
|
indicates run time remaining (timed run) or elapsed time (hold run)
|
Temperature setting range
|
-20 to +40°C (in 1°C increments)
|
Rotor temperature
|
±2°C of set temperature (after equilibration)
|
Ambient temperature range
|
16 to 38°C (60 to 100°F)
|
Acceleration
|
Maximum or slow
|
Deceleration
|
Maximum, slow, or off
|
Power rating
|
200/208/240V, 50/60Hz, 24A
|
Dimensions (W x D x H)
|
28 x 33 x 45.5 in
|
Weight
|
583 lb (264.5 kg)
|