- Beckman Coulter Avanti J-26 XPI High-Speed Refrigerated Floor Centrifuge 208V (393127)
- Attached 208V Power cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Avanti J-26 XPI refrigerated floor centrifuge with the following results:
- Ran to 20,000 rpm: PASSED
- Refrigeration held at -9°C: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no issue found: PASSED
Manufacturer
|
Beckman Coulter
|
Model
|
Avanti J-26 XPI
|
Catalog No.
|
393127
|
Serial No.
|
JXT12G01
|
Mfg. date
|
July 26, 2012
|
Set speed range
|
100 to 26,000 rpm (in 100-rpm increments below 10,000 rpm, 500-rpm increments above 10,000 rpm) or equivalent RCF (in 100 x g increments)
|
Elutriation speed setting range
|
0 to 5,000 rpm (in 10-rpm or 10 x g increments)
|
Time setting range
|
1 min to 99 hr 59 min or continuous (hold)
|
Time display
|
indicates run time remaining (timed run) or elapsed time (hold run)
|
Set temperature range
|
-10 to +40°C (in 1°C increments)
|
Temperature accuracy
|
rotor temperature controlled to within +/-2°C of set temperature (after equilibration)
|
Ambient temperature range
|
16 to 36°C (60 to 100°F)
|
Acceleration
|
maximum or slow
|
Deceleration
|
maximum, slow, or off
|
Power rating
|
200/208/240V, 50/60Hz, 24A
|
Dimensions (W x D x H)
|
28 x 34 x 45.5 in
|
Weight
|
640 lb (290 kg)
|