- Tecan Infinite F Plex 200 Pro Multi-Mode Microplate Reader (30050303)
- USB Cable
- Power Cable
- 90-Day Warranty
- FP 485(20) - Pol
- FP 485(20) Pol
- FI 620(20)
- ABS 560(0)
- 535(25)
- 535(25)
- 680(30)
After its initial disinfection and inspection, our technicians tested the Tecan Infinite F Plex 200 Pro microplate reader with the following results:
- Power and initialization: PASSED
- Plate loading drawer: PASSED
- Ran with standard ABS plate; consistent and stable readings: PASSED
- Shaking: PASSED
Notice: This reader has some minor cosmetic blemishes. Please review the available photos to see the physical condition before purchasing.
|
Manufacturer
|
Tecan
|
|
Model
|
Infinite 200 Pro
|
|
Configuration
|
Infinite F Plex
|
|
Reference No.
|
30050303
|
|
Serial No.
|
1904000172
|
|
Mfg. date
|
20190401
|
|
Power rating
|
100-120VAC / 220-240VAC, 50/60Hz
|
|
Power consumption
|
150VA
|
|
Dimensions (W x D x H)
|
17 x 18 x 9.5 in
|
|
Weight
|
32 lb
|
|
Light source
|
UV Xenon flashlamp
|
|
Bandwidth
|
Ex: < 5 nm for λ ≤ 315 nm and < 9 nm for λ > 315 nm; Em: < 20 nm
|
|
Wavelength accuracy
|
Absorbance: < ± 0.5 nm for λ > 315 nm; < ± 0.3 nm for λ ≤ 315 nm Fluorescence: < ± 2 nm for λ > 315 nm; < ± 1 nm for λ ≤ 315 nm
|
|
Wavelength reproducibility
|
Absorbance: < ± 0.5 nm for λ > 315 nm; < ± 0.3 nm for λ ≤ 315 nm Fluorescence: < ± 1 nm for λ > 315 nm; < ± 0.5 nm for λ ≤ 315 nm
|
|
Infinite F configurations
|
Up to 4 filter pairs per slide
|
|
Wavelength range
|
Fluorescence intensity: Ex 230 – 850 nm, Em 280 – 850 nm Absorbance: 230 - 1000 nm
|
|
Detectors
|
Fluorescence – PMT, UV and red-sensitive Absorbance – UV silicon photodiode Luminescence –photon counting system with low dark current PMT
|
|
Plate formats
|
6- to 384-well plates, cuvettes, NanoQuant plate
|
|
Temperature control
|
Ambient +5°C up to 42°C
|
|
Shaking
|
Linear, orbital
|
|
Fluorescence sensitivity values
|
Top reading: 85 amol / well (100 μl; 384-well plate) Bottom reading: 5.0 fmol / well (200 μl; 96-well plate) TRF: 2.8 amol / well (100 μl; 384-well plate) FP: <4 mP standard deviation @ 1 nM Fluorescein
|
|
Luminescence sensitivity values
|
Glow: 225 amol ATP / well (25 μl; low volume 384-well plate) Flash: 12 amol ATP / well (55 μl; 384-well plate)
|
|
Absorbance ratio accuracy @ 260 / 280 nm
|
± 0.07
|
|
Absorbance precision @ 260 nm
|
< 0.2%
|
|
Absorbance accuracy @ 260 nm
|
< 0.5%
|
|
Absorbance measurement range
|
0 - 4 OD
|
|
Injector pump speed
|
100 - 300µl/s
|
|
Injector injection volume
|
selectable in 1µl increments; max. volume: 800 µl per stroke
|
|
Injector dead volume
|
100 µl including pump back
|
|
Fastest read times
|