- New Brunswick Scientific Innova 40 Orbital Incubator Shaker 120V (M1299-0010)
- Shaker Platform
- Power Cable
- 90-Day Warranty
Notice: We have flask clamps available for an additional cost. Please contact us if you need flask clamps added to your purchase.
After its initial disinfection and inspection, our technicians tested the New Brunswick Scientific Innova 40 orbital incubator shaker with the following results:
- Shaking holds steady at setpoint: PASSED
- Shaking calibration: PASSED
- Temperature holds steady at 37°C setpoint: PASSED
- Temperature calibration: PASSED
- Installed a new belt.
- Installed new gas springs.
Notice: This shaker has some minor cosmetic blemishes. Please review the available photos to see the physical condition before purchasing.
|
Manufacturer
|
New Brunswick Scientific
|
|
Model
|
Innova 40
|
|
Cat. No.
|
M1299-0010
|
|
Serial No.
|
700423789
|
|
Mfg. date
|
2007
|
|
Orbit
|
1 in (2.5 cm)
|
|
Temperature range
|
Ambient +5°C to 80°C
|
|
Temperature accuracy and control
|
±0.1°C at 37°C with a nominal uniformity of ±0.25°C
|
|
Refrigerated
|
No
|
|
Shaking speed range
|
25 - 500 rpm (depending on load)
|
|
Shaking speed control
|
±1 rpm
|
|
Drive mechanism
|
Triple-eccentric counterbalanced drive in cast iron housing with 9 permanently lubricated ball bearings. Driven by solid-state brushless DC motor
|
|
Timer and parameter control
|
Speed and temperature continuous. Shaker can be set to run from 0.1 to 99.9 hours. On-board firmware can program multiple parameter changes to temperature and speed. Shaker control and data logging via RS-232 and remote computer running conventional laboratory management software or HyperTerminal.
|
|
Alarms
|
Indicate when speed deviates more than 5 rpm or temperature more than 1°C from setpoint, and when program or time has finished. Alarm may be muted.
|
|
Safety
|
Acceleration circuit prevents sudden starts and stops. Sensor shuts off temperature when high-temperature limit is exceeded.
|
|
Computer interface (RS-232)
|
For data logging and control using laboratory management software or HyperTerminal.
|
|
Setpoint retention
|
Setpoints and operating status retained by non-volatile memory. Automatic restart after power is restored (indicated by flashing display)
|
|
Ambient conditions
|
10°C to 35°C. 20 to 80% RH (non-condensing)
|
|
Power rating
|
120V, 60Hz, 1Ph
|
|
Power consumption
|
800VA
|
|
Platform dimensions (W x D)
|
18 x 18 in (46 x 46 cm)
|
|
Exterior dimensions (W x D x H)
|
22 x 30 x 24 in (55.9 x 76.2 x 61 cm)
|
|
Weight
|
138 lb (62.7 kg)
|