- Apricot Designs I-Pipette Pro 96-500 Automated Multichannel Pipetting System
After its initial disinfection and inspection, our technicians tested this I-Pipette Pro 96-500 with the following results:
- Simulates loading tips: PASSED
- Performs aspiration: PASSED
- Mixes dispense in both manual and automatic mode: PASSED
- Lifts plate carrier to tips aspirated, mixed and dispensed: PASSED
Note: The data cable to the PDA has been replaced. The cable connector was also re-nutted.
|
|
|
|
|
|
|
(96-500) <3%CV at 5ul (96-125/384) <3%CV at 1ul
|
|
(96-500) +/-2% at 10ul (96-125/384) +/- 2% at 1uL
|
|
(96-500) 1ul (96-125/384) 0.1ul
|
|
(96-500) 1-500 ul (96-125/384) 0.5-125 ul
|
|
|
|
|
|
|
|
|