Beckman Coulter Biomek NXp Automated Liquid Handling Workstation
 
Successful Software Communication & Plate Movement 
 
After connecting the NXp to the Biomek software, we performed the following tests:
 
    - NXp initializes and communicates with Biomek software: PASSED
    - Home all of the motors: PASSED
    - Prime syringe pumps: PASSED
    - Aspirate/dispense fluids: PASSED
    - Load/unload tips: PASSED
    - Lift/move microplates: PASSED
    - Create/edit/remove programs: PASSED
    - Beckman Coulter Biomek NXp Automated Liquid Handler
    - MultiChannel 96 P20/200 Head with Gripper
    - PC with software installed
    - Monitor
    - Mouse, keyboard
    - Tip loader
    - Heating block with controller
    - Accuframe sensor
    - Accuframe tool
    - Metal reservoir with adapter plate
    - Serial cable
    - Power cables
Consumables are required for this system to be used in production. You will need to acquire the consumables separately. They can generally be purchased from the manufacturer or a third-party dealer.
 
| Manufacturer | Beckman Coulter | 
| Model | Biomek NXp | 
| Catalog No. | A31841 | 
| Serial No. | A318410148 | 
| Environment | Indoor use only | 
| Compressed Air RequirementsNX-MC: Requires compressed air NX-S8: Only Span-8 instruments that use a Microplate Shaking ALP require compressed air
 
 | Air Pressure: 60 - 100 psi of compressed air414 - 690 kPa of compressed air
 Air Flow: 2.25 cfm @ 60 psi 0.064 m3/min @ 414 kPa
 | 
| System Fluid RequirementsNX-S8: Requires system fluid 
 | De-ionized or Distilled waterSystem fluid ambient operating temperature is 59°–86°F (15°–30°C).
 System fluid should be degassed for 24 hours prior to use.
 | 
| Weight | 190 lbs. (86.2 kg) | 
| Ambient Operating Temperature | 41°-86°F (5°-30°C) | 
| Humidity Restrictions | <85% (non-condensing) @ 86°F (30°C) | 
| Altitude Restrictions | up to 6562 ft (2000 m) | 
| Installation Category | Category II | 
| Pollution Degree | 2 | 
| Sound Pressure Level | Maximum sound pressure:  82 dBMaximum sound pressure at 1 meter away: 75 dB
 | 
| Fuses |  US: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, UL recognized, CSA certifiedEurope: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, CENELEC approved
 | 
| Communications to Host | RS-232 port | 
| Communications to Tip LoaderNX-MC: Uses a Tip Loader
 | CAN | 
| Communications to Active ALPs via the Biomex NXp Instrument | CAN | 
| Power Requirements | 100-240V, 4A, 50/60Hz | 
| Overall Dimensions (WxDxH) | 35 x 31 x 42" |