- Beckman Coulter Optima L-80 XP Preparative Floor Ultracentrifuge (392051)
- x4 Feet
- Attached Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Optima L-80 XP floor ultracentrifuge with the following results:
- Ran to 80,000 rpm: PASSED
- Refrigeration reached 0°C when idle: PASSED
- Vacuum reached below 20 microns: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no issue found: PASSED
- Changed vacuum and diffusion oil.
- Changed computer bios battery.
- Replaced bellow seals.
- Installed refurbished tach/hall effect sensor.
Manufacturer
|
Beckman Coulter
|
Model
|
L-80 XP
|
Catalog No.
|
392051
|
Serial No.
|
LXP08M24
|
Mfg. date
|
2008
|
Speed
|
1,000 to 80,000 rpm in 100-rpm increments
|
Speed Control
|
Actual rotor speed, ±10 rpm of set speed after equilibration (above 1 000 rpm)
|
Speed Display
|
Actual rotor speed in 10-rpm increments below 1 000 rpm and 100-rpm increments ≥1 000 rpm
|
Set Time
|
To 999 hours 59 minutes; Hold for continuous runs
|
Time Display (Manual Operation)
|
Indicates time remaining in timed runs, time elapsed in Hold runs, or estimated time remaining in ω2t runs
|
Time Display (Programmed Operation)
|
Time remaining in step
|
Set Temperature
|
0 to 40°C in 1°C increments
|
Temperature Control (After Equilibration)
|
±0.3°C of set temperature
|
Temperature Display
|
Actual rotor temperature in 0.1°C increments
|
Ambient Temperature Range
|
10 to 40°C
|
Humidity Restrictions
|
<95% (noncondensing)
|
Vacuum
|
Below 5 microns (0.7 Pa)
|
Acceleration
|
10 acceleration profiles— 9 slow rates or maximum acceleration from 0 to set speed
|
Deceleration
|
11 deceleration profiles—9 slow rates, coasting to a stop without brake, or full dynamic braking from set speed
|
Power rating
|
200/240V, 50/60Hz, 20A
|
Dimensions (W x D x H)
|
37 x 26.5 x 49.5 in
|
Weight
|
1025 lb (465 kg)
|