- Beckman Coulter PA 800 plus Pharmaceutical Analysis System (A74603)
- x2 36-Vial Buffer Trays
- National Instruments GPIB-USB-HS Controller Cable
- Power Cable
- 90-Day Warranty
- PC Workstation and Control Software
Notice: A PC workstation and control software must be acquired separately to operate this system.
After its initial disinfection and inspection, our technicians tested the Beckman Coulter PA 800 plus Pharmaceutical Analysis System with the following results:
- Power and initialization: PASSED
- Power LED indicator turns on: PASSED
- UV LED indicator turns on: PASSED
- Trays homed: PASSED
- Performed inspection; no issue found: PASSED
Notice: This system has some very minor cosmetic blemishes. Please review the available photos to see the physical condition before purchasing.
|
Manufacturer
|
Beckman Coulter
|
|
Model
|
PA 800 plus
|
|
Cat. No.
|
A74603
|
|
Serial No.
|
A746030784
|
|
Mfg. date
|
July 03, 2013
|
|
Power rating
|
100-240V, 50/60Hz, 8A
|
|
Power consumption
|
Supply voltage must not exceed 10% of nominal
|
|
Fuses
|
8.0 A slow blow, ¼ in. (2 each), 100 to 120 VAC 6.3 A time delay, 20 mm (2 each), 200 to 240 VAC
|
|
Installation (overvoltage) category
|
Category II
|
|
Maximum heat dissipation
|
400 W (1,024 BTU/hr)
|
|
Pollution degree
|
2
|
|
I/O
|
TTL: 2 Contact closures: 2
|
|
Working environment
|
Altitude: Up to ≤ 2,000 m (6,562 ft) Humidity: < 80% (noncondensing) at 15°C to 30°C (59°F to 86°F) < 60% (noncondensing) at 30°C to 40°C (86°F to 104°F) Temperature: 15°C to 40°C (59°F to 104°F), 15°C to 30°C (59°F to 86°F) recommended
|
|
Dimensions (W x D x H)
|
25 x 28.5 x 29 in (63.5 x 72.4 x 73.7 cm)
|
|
Weight
|