- Beckman Coulter Vi-CELL XR Cell Viability Analyzer (383556)
- Computer with Vi-CELL XR 2.04 Software Installed
- Monitor, Keyboard, and Mouse
- Communication Cables
- Power Cables
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Vi-CELL XR cell viability analyzer with the following results:
- Powers on and initializes; connects to control software: PASSED
- Reagent pack error occurred as expected: PASSED
- Flushed lines with water to test reagent sensor: PASSED
- Pumps checked: PASSED
- Camera image checked: PASSED
Notice: The flow cell has been protected with a piece of cardboard to prevent the camera from smashing the cell during transit. It will need to be removed before use.
Manufacturer
|
Beckman Coulter
|
Model
|
Vi-CELL XR
|
Reference No.
|
383556
|
Serial No.
|
AW09236
|
Mfg. date
|
2014/02/28
|
Autosampler
|
Yes
|
Size range
|
2 - 70 um
|
Sample volume
|
500 uL
|
Analysis time
|
2.5 min
|
Imaging technology
|
Camera focus: auto Image collection: Firewire Camera: 1394x1040 CCD
|
Image zoom
|
Yes
|
Aspiration and Trypan Blue mixing
|
Variable
|
Concentration range
|
5 x 104 to 1 x 107 cells/mL Counting accuracy: +/- 6%
|
Operating system
|
Windows 7 or Windows 10
|
Instrument type
|
Video imaging through a quartz flow cell
|
Temperature
|
10° to 40°C (50°F to 104°F)
|
Power rating
|
100/240V, 50/60Hz, 0.5/1A
|
Power consumption
|
120W
|
Dimensions (W x D x H)
|
15 x 17.5 x 17.5 in
|
Weight
|
36.2 lb (16.4 kg)
|