- BioTek Instruments ELx405 Select Microplate Washer (ELX405UVS)
- 96-tube Dual-Action Manifold
- Vacuum Pump
- Buffer Bottle
- x2 Waste Bottles
- Tubing
- Pump Power Cable
- Main Power Cable
- 90-Day Warranty
Notice: The included 96-tube Dual-Action manifold can wash both 96- and 384-well plates without hardware changes.
After its initial disinfection and inspection, our technicians tested the BioTek Instruments ELx405 Select microplate washer with the following results:
- Power and initialization: PASSED
- Priming: PASSED
- Dispensing: PASSED
- Aspiration: PASSED
- Soaking: PASSED
Notice: This microplate washer has some cosmetic blemishes such as the scratch on the control panel and wear on the drip tray. Please review the available photos to see the physical condition before purchasing.
|
Manufacturer
|
BioTek Instruments
|
|
Model
|
ELx405 Select
|
|
Model No.
|
ELX405UVS
|
|
Serial No.
|
241611
|
|
Microplate type
|
96- and 384-well Low profile and standard height
|
|
Onboard software
|
Create, edit and run multiple protocols
|
|
Soaking
|
Programmable in minutes and seconds, up to 60 min.
|
|
Ultrasonic advantage
|
Yes (option)
|
|
Manifold type
|
96-tube Dual Action manifold for 96- & 384-well washing
|
|
Volume range
|
50 to 3000 microliter increments
|
|
Wash cycles
|
1-10
|
|
Buffer/reagent selection
|
Auto switching module for up to 4 buffers (option)
|
|
Supply bottle
|
4 or 10 L (optional)
|
|
Dispense precision
|
≤3% CV: 300 µL/well (96-well washing)
|
|
Residual volume
|
≤2 µL/well (standard 96-well plates)
|
|
Wash speed
|
96-wells: 300 μL/well, 3 cycles: <30 s 384-wells: 100 μL/well, 3 cycles: <80 s
|
|
Flow rates
|
High flow to low flow Optimized rates for cell assays
|
|
Sterilization
|
Chemical
|
|
Vacuum filtration*
|
Selectable: approximate final at 30 s: Lowest: -38 mmHg Low: -113 mmHg Medium: -200 mmHg High: -390 mmHg Highest: -506 mmHg Vacuum filtration time range: 5-999s
|
|
Interface
|
RS-232 port
|
|
Power rating
|
100 to 240V~, 50/60Hz, 8.0A
|
|
Accessory outlet
|
≤5.0A, used for vacuum pump
|
|
Dimensions (W x D x H)
|
14 x 17 x 10 in (35.6 x 43.2 x 25.4 cm)
|
|
Weight
|
With internal buffer switching: 36 lb (16.5 kg) Without internal buffer switching: 30 lb (13.5 kg)
|
*With available accessory module sold separately.
Lower Costs: Save 30%–70% off standard retail prices on fully tested, pre-owned instruments.
Quality Equipment: Work with confidence knowing every instrument undergoes thorough testing for real-world lab environments.
Faster Delivery: Minimize downtime with in-stock inventory ready to ship immediately following a final quality inspection.
Dedicated Support: Experience customer service with guaranteed responses within one business day and full after-sale support.
Warranty Options: Purchase securely with warranty protection included on all equipment.