- Covaris LE220-plus Focused-Ultrasonicator
- Laptop with SonoLab 8 Software Installed
- USB Cable
- Laptop Power Adapter
- Power Cable
- 90-Day Warranty
Notice: Accessories and consumables must be acquired separately to operate this instrument.
After its initial disinfection and inspection, our technicians tested the Covaris LE220-plus focused-ultrasonicator with the following results:
- Instrument and software connection: PASSED
- Initialization and homing: PASSED
- Degas: PASSED
- Door opening: PASSED
- Sonicator performance: PASSED
- Moved into service position: PASSED
|
Manufacturer
|
Covaris
|
|
Model
|
LE220-plus
|
|
Serial No.
|
004248
|
|
Treatment power
|
2.5 to 500 Watts Peak Incident Power 0.1 to 250 Watts Average Incident Power
|
|
Operating system
|
Includes: Notebook computer interface via USB with Microsoft Windows and Covaris SonoLab Operating Software installed
|
|
Chiller
|
ThermoCube and WCS 2.0 required
|
|
Sample processing capacity
|
1 to 8 tubes at a time parallel processing; 96 tube capacity
|
|
NGS sample volume
|
5 to 500 µL
|
|
Sample volumes
|
truXTRAC® FFPE = LCM, 5 to 20 micron slides, up to 80 micron slices, and 1.2 mm cores truXTRAC cfDNA = up to 4 ml plasma truXTRAC DBS (dried blood spots) = up to seven 3 mm punches truCHIP® (chromatin) = up to 1 mL for mammalian cells truCOLLECT™ = up to 35 µL fresh whole blood
|
|
Recommended batch size
|
96+ samples
|
|
Covaris consumables
|
Individual microTUBEs; 8 microTUBE strips (rack required); 96 microTUBE plate; 96 oneTUBE-10 AFA plate, miniTUBEs, milliTUBEs
|
|
Upgrade potential
|
Robotics upgrade capable (PN 500578)
|
|
Operator skill requirement
|
Minimal
|
|
Automatic instrument set-up
|
Yes
|
|
Integration with lab automation
|
Yes
|
|
Operating environment
|
Ambient temperature: 19°C to 25°C (66°F to 77°F) Relative humidity: 30% to 70%
|
|
Regulatory labeling
|
CE, ETL Mark (for Product Safety), WEEE
|
|
Safety
|
Complies with Low Voltage Directive 2006/95/EC. Certified to IEC/EN/ANSI/UL 61010-1:2010 and CAN/CSA C22.2 No. 61010-1, “Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 1: General Requirements”
|
|
EMC
|
Complies with Class A Industrial/Scientific/Medical (ISM) equipment under EN 61326-1 for EU EMC Directive 2014/30/EU. Also FCC Part 15 Class A radio emissions requirements for the USA and ICES-003 Class A for Industry Canada
|
|
Power consumption
|
500VA max.
|
|
Power rating
|
100-240V~, 50/60Hz
|
|
Dimensions (W x D x H)
|
24 x 30 x 19 in (61 x 76 x 48 cm)
|
|
Weight
|
approx. 110 lb (50 kg)
|