**Currently en route to our facility. It will take about 30 days to arrive, after which we will test and certify it.**
Included Items:
- SP Scientific VirTis Benchmark 3000 Lyophilizer W/ Lyos Control Software
- Power Cable
Certified to Perform at Manufacturer Specifications
After your instrument arrives at our facility, it will be inspected and tested to certify it performs at manufacturer specifications. If any component does not pass testing, then it will receive the service necessary to restore full functionality.
Specifications
|
Manufacturer
|
SP Scientific
|
|
Model
|
Benchmark 3000
|
|
Lowest shelf temperature (shelf freezing)
|
≤ -60°C
|
|
Maximum shelf temperature
|
≥ 65°C
|
|
Shelf inlet temperature control range and accuracy, concurrent with condenser operation (drying mode)
|
-55°C to 65°C (±1°C)
|
|
Shelf pull-down from 20°C to -40°C
|
≤ 40 minutes
|
|
Shelf heat-up rate from -40°C to 20°C
|
≥ 1°C / minute
|
|
Condenser blank-off temperature (Lowest condenser temperature)
|
≤ -75°C
|
|
Condenser pull-down from 20°C to -45°C (no shelf assist mode)
|
≤ 30 minutes
|
|
Maximum temperature deviation across the top of the product shelf at steady state conditions from high to low within a temperature range of -40°C to 40°C
|
≤ ± 1°C
|
|
Maximum temperature deviation between any two similarly located points on any two product shelves at steady conditions within a temperature range of -40°C to 40°C from high to low
|
≤ ± 1°C
|
|
System evacuation time to 100 millitorr (with the primary vacuum pump and without vacuum line or exhaust filters
|
≤ 30 minutes
|
|
System blank-off pressure (lowest system vacuum)
|
≤ 15 millitorr
|
|
Vacuum level control and range (Pressure tolerance when lowering and maintaining chamber pressure during drying cycles)
|
50 to 500 millitorr (±5 mT)
|
|
Condenser Capacity (minimum)
|
≥ 42 liters/24 hours
|
|
Defrost Time
|
≤ 60 minutes
|
|
Shelf Nominal Dimensions (W x D x H)
|
24 x 36 x 0.5 in
|