- Tuttnauer 2340M Manual Autoclave Steam Sterilizer (62683)
- Tray Rack
- x2 Trays
- Power Cable
- 90-Day Warranty
Notice: This autoclave comes with 2 trays. Both trays have the following overall dimensions (W x D x H): 7.6 x 17.5 x 0.8”
After its initial disinfection and inspection, our technicians tested the Tuttnauer 2340M manual autoclave steam sterilizer with the following results:
- Ran all several cycles without error: PASSED
- Fluid filling and draining: PASSED
- Chamber heating: PASSED
- Chamber pressure: PASSED
- Door switch: PASSED
- Door seal checked for leaks: PASSED
- Checked for internal leaks: PASSED
- Pilot lights: PASSED
- Pressure gauge: PASSED
- Replaced 3-way valve.
- Replaced safety valve.
- Replaced pilot lights.
- Replaced front stickers.
- Painted panels.
- Cleaned reservoir.
|
Manufacturer
|
Tuttnauer
|
|
Model
|
2340M
|
|
N. B. No.
|
62683
|
|
Serial No.
|
9704977
|
|
Mfg. date
|
1997
|
|
Chamber volume
|
5 gal (19 liters)
|
|
Max. pressure
|
2.76 bar
|
|
Test pressure
|
4.20 bar
|
|
Standard cassette capacity
|
2 full & 2 half
|
|
Standard unwrapped cycle time
|
Cold: 27 mins Hot: 13 mins
|
|
No. of trays
|
3
|
|
Standard tray dimensions (W x D x H)
|
6.7 x 16.3 x 0.8 in (17 x 41.5 x 2 cm)
|
|
Chamber dimensions (Dia. x D)
|
9 x 18 in (23 x 47 cm)
|
|
Power consumption
|
1400W
|
|
Power rating
|
120V~, 50/60Hz, 11.7A, 1Ph
|
|
Dimensions (W x D x H)
|
20 x 21.5 x 14.4 in (51 x 54.5 x 36.5 cm)
|
|
Weight
|
79 lb (36 kg)
|